LC-80 ChroMini HPLC Liquid Chromatograph

Flow rate range: 0.001-5.000 mL/min, in 0.001 mL increments

Flow accuracy: ≤±0.1% (1.00ml/min, H2O,R.T.,8.5MPa);

Flow precision: RSD≤0.5%  (1.00ml/min, H2O,R.T.,8.5MPa)

Maximum pressure: 5800 psi;

Pressure accuracy: ≤±1.0%;

Pressure pulsation: ≤0.1 MPa

P80 C  Dual-piston reciprocating parallel   

Flow Presion: RSD<0.5%;

Fluctuation: 0.1MPa ;

Mixing Accuracy :±1%;

Mixing Precision:±0.2%;

Power Supply:   110V / 220V

UV80 Detector  

Wavelength Range: 190-680 nm

Flow cell volume: 8nm;

Wavelength Accuracy: ±1 nm;

Wavelength Precision: <0.1nm;

Minimum Detectable Concentration: 5×10-9 g/mL;

Noise :±0.25×10-5AU(Static)/±1×10-5AU(dynamic, under specified requirements);

Drift :≤0.4×10-4AU/h(Static)/±2×10-4AU/h (dynamic, under specified requirements);

Power Supply: 110V/220V;

WS-80 Chromatographic Data-processing Workstation

Signal Transmission Form: Full digital signal;

Operation Mode: Reflecting and monitoring the operation of instruments’ each unit (editing gradient program for changing wavelength scanning, setting event procedure, spectrum scanning, spectrogram comparison, etc.)

COLUNM OVEN Original Series C18  5μm ID4.6×250mm

Enquiry For LC-80 ChroMini HPLC Liquid Chromatograph