Wuhan Bonnin Technology Ltd, Laboratory Instruments Expert in China

Product Description

DSC-300C Low Temperature Differential Scanning Calorimeter is designed for the measurement of DSC thermal analysis in plastics and rubber industry, can be glass transition temperature test, phase transition test, melting and enthalpy test, product stability, oxidation induction test, etc. Low temperature differential scanning calorimeter DSC300C, liquid nitrogen refrigeration, high sensitivity.

Test standards:

ISO/TR10837:1991, ASTM D3895-1998, ASTM E 967, ASTM E 968, ASTM E 793, ASTM D 3895, ASTM D 3417, ASTM D 3418, ISO 11357-6

Main features:

1. Brand new furnace structure, better resolution and resolution and better baseline stability

2. Parameters can be set to multi-stage heating, constant temperature, cooling

3. The instrument can adopt two-way control (host control, software control), with friendly interface and easy operation

4. Using imported chips, acquisition circuit shielding and anti-interference processing

5. The sensor adopts hot-melt technology instead of traditional spot welding technology, with higher sensitivity

Advantages:

1. The industrial-grade 7-inch touch screen of the DSC300C differential scanning calorimeter displays rich information.
2. DSC300C differential scanning calorimeter has a new metal furnace structure with better baseline and higher precision.
3. USB communication interface, strong versatility, reliable and uninterrupted communication, supports self-recovery connection function.
4. Automatic switching of two air flow rates, fast switching speed and short stabilization time. At the same time, add a protective gas input.
5. The DSC300C differential scanning calorimeter adopts imported chips with high stability.

Technical datas:

Model DSC-300C DSC-300L
Temperature range -40~600℃ -170~600℃
Temperature resolution 0.01°C 0.001℃
Temperature fluctuation ±0.01℃ ±0.01℃
Heating rate 0.1~100℃/min 0.1~100℃/min
Cooling rate 0.1~40℃/min 0.1~40℃/min
DSC range 0~±600mW 0~±600mW
DSC resolution 0.01uW 0.01uW
DSC sensitivity 0.001mW 0.001mW
Temperature control mode heating, constant temperature, cooling (automatic program control), constant temperature time can be set by yourself
Scanning mode Heating scan, cooling scan
Atmosphere control gas Two-way automatic switching (instrument automatic switching)
Display mode 24bit color, 7-inch LCD touch screen display
Program control can realize six-stage heating constant temperature control, special parameters can be customized
Power supply AC220V/50Hz or 110V, 60Hz
Gas flow 0-300mL/min
Gas pressure ≤5MPa
Data interface Standard USB interface
Parameter calibration Equipped with standard substances (indium, tin, lead), users can correct the temperature

Enquiry For DSC-300C Low Temperature Differential Scanning Calorimeter