Wuhan Bonnin Technology Ltd, Laboratory Instruments Expert in China

Product Description

Auto-Pure 20 series nucleic acid purification system has further improved the sample processing capacity based on 32A. The maximum processing volume of 20A is 3 mL; and the maximum processing volume of 20B/24D is 5 mL and 10 mL, combined with the patented left and right mixing method and reagent strips. In addition, the single-strip design of the kit can effectively avoid the waste of the kit, which is more suitable for customers with a small amount of samples.

AUTO-PURE 20 AUTOMATIC NUCLEIC ACID DNA EXTRACTION SYSTEM

 

 

 

 

 

 

 

Features:

•Large sample volume processing: 20B maximum processing volume 5 mL, 24D maximum processing volume 10 mL, to meet a variety of applications

•24D has two maximum processing volumes, suitable for 5 mL and 10 mL consumables

•Patented designed kits can be individually packaged for nucleic acid extraction and purification from a single sample

•The patented designed nucleic acid sample left and right mixing method is especially suitable for cfDNA extraction, nucleic acid sample library construction, blood nucleic acid screening, and nucleic acid hypersensitivity detection.

•Drawer design for easy kit placement

•Open instrument parameter settings, can set freely: mixing speed, magnetic absorption method, heating method and other functions

•24D operation time, 15-30 min/time

 

Technical datas:

Model Auto-Pure24D
Throughput 1~24
Process volume 50~10000 µL/50~5000 µL
Consumable 5 mL tube strips+magnetic rod’s tip/10 mL tube strips+magnetic rod’s tip
Principle/working mode Magnetic bead method, magnetic rod method/up and down mixing, left and right mixing
Purification accuracy 100 copy sample positive rate>95 %
Stability CV<5 %
Collection efficiency >95 %
Heating for lysis tube Ambient temperature~120 °C
Heating for elution tube Ambient temperature~120 °C
Operation 7-inch touch screen, 3 shortcut keys, external mouse
Internal program Preset 8 programs, max store 100 programs
Protocol management Create, edit, delete, protocol mode
Extension interface 4 standard USB port, Ethernet port, wifi, built-in SD card
Lighting Yes
Pollution control UV sterilization
Exhaust Fan
Power supply 450 W
Dimension 400×520×450 mm
Weight 30 kg

Enquiry For Auto-Pure 24D Nucleic Acid Extraction Purification System